4 Command and Query Summary
K0472 Revision A 4-88
:SOUR:PRES:LEV:IMM:AMPL
Applicability: PACE5000, PACE6000
Command Syntax
:SOURce[x][:PRESsure][:LEVel][:IMMediate][:AMPLitude] <number>
where x = 1 or 2 is the module number
(default - 1)
Parameter: <decimal> Pressure in current units
Short form: SOUR <number>
Function: Set the pressure set-point
Default: 0.0
Query Syntax
:SOURce[x][:PRESsure][:LEVel][:IMMediate][:AMPLitude]?
where x = 1 or 2 is the module number
(default - 1)
Short form :SOUR?
Function: Programmable set-point value
Response: Decimal number representing pressure set-point in current units.
Description
This command sets the pressure set-point and is the long form of :SOUR:PRES.
Example
TX> :SOUR:PRES?
RX> :SOUR:PRES:LEV:IMM:AMPL 0.4000000
The current controller set-point is 0.4 of the current unit.
It can be set to another value, such as:
TX> :SOUR:PRES 0.5
TX> :SOUR:PRES?
RX> :SOUR:PRES:LEV:IMM:AMPL 0.5000000