PACE Series SCPI Manual
K0472 Revision A 4-75
:SENS:SPE:INL
Applicability: PACE6000
Command Syntax
:SENSe[x][:SPEed]:INLimits <number>
where x = 1 or 2 is the module number
(default - 1)
Parameter: <decimal> airspeed in limits value as % full-scale
Short form: :SENS:SPE:INL <number>
Function: The instrument has a airspeed in-limits indicator. This can generate a
service request when the airspeed value is within limits for a set time
period.
Default: 0.01 % full-scale
minimum 0 % full-scale
maximum 100 % full-scale
Query Syntax
:SENSe[x][:SPEed]:INLimits?
where x = 1 or 2 is the module number
(default - 1)
Short form: :SENS:SPE:INL?
Function: Query airspeed in-limits value
Response: Decimal representing airspeed in-limits value as % full-scale.
Description
Sets the in-limits value.
Note: If the instrument is not in aero mode it records an error: -221. Settings conflict.