PACE Series SCPI Manual
K0472 Revision A 4-107
:STAT:QUES:ENAB
Applicability: PACE1000, PACE5000, PACE6000
Command Syntax
:STATus:QUEStionable:ENABle <integer>
Parameter: <integer> 16 bit value to set enable bits
Short form: STAT:OPER:ENAB <integer>
Function: Controls the questionable data condition enable register.
Default: 0
minimum 0
maximum 32767
Query Syntax
:STATus:QUEStionable:ENABle?
Short form: :STAT:OPER:ENAB?
Function: Query questionable data condition enable register.
Response: 16 bit value of enable register.
Description
Controls the bits that pass through the questionable data condition reporting system, see
status reporting section.