4 Command and Query Summary
K0472 Revision A 4-48
SENS:MACH:SLEW
Applicability: PACE1000, PACE6000
Command Syntax
SENSe[2]:MACH:SLEW
where x = 1 or 2 is the module number
(default - 1)
Parameter:
Short form: SENS:MACH:SLEW
Function: Sets the rate set-point for MACH slew.
Query Syntax
SENSe:MACH:SLEW?
Short form: SENS:MACH:SLEW?
Function: Queries the rate set-point for MACH slew.
Response: A decimal, pressure reading/second in MACH.
Description
This command sets the rate set-point for MACH slew. The query gets the rate set-point for
MACH.
Note: If the instrument is not in aero mode it returns an error: -221. Settings conflict.