EasyManua.ls Logo

GE PACE5000

GE PACE5000
176 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4 Command and Query Summary
K0472 Revision A 4-92
:SOUR:PRES:RANG:LOW
Applicability: PACE5000, PACE6000
Command Syntax
n/a
Parameter:
Short form:
Function:
Default:
Query Syntax
:SOURce[x]:PRESsure:RANGe:LOW?
where x = 1 or 2 is the module number
(default - 1)
Short form: :SOUR:LOW?
Function: Queries the lower full-scale of the module [x].
Response: Decimal number in current selected units.
Description
This query gets the lower full-scale of the module [x].
Example 1
TX> :Sour:Pres:Rang:Low?
RX> :SOUR:PRES:RANG:LOW "-1000mbar g"
Example
2
TX> :Sour1:Pres:Rang:Low?
RX> :SOUR:PRES:RANG:LOW "-1000mbar g"

Table of Contents

Other manuals for GE PACE5000

Related product manuals