EasyManuals Logo

GE PACE5000 Technical Manual

GE PACE5000
176 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #121 background imageLoading...
Page #121 background image
PACE Series SCPI Manual
K0472 Revision A 4-93
:SOUR:PRES:SLEW
Applicability: PACE5000, PACE6000
Command Syntax
:SOURCE[x][:PRESsure]:SLEW <number>
where x = 1 or 2 is the module number
(default - 1)
Parameter: <decimal> rate in pressure units/second or enumerated string maximum
or minimum rate.
Short form: SOUR:SLEW <number> or maximum or minimum rate
Function: Selects the pressure rate used when value rate is selected.
Default: 100
Query Syntax
:SOURCE[x][:PRESsure]:SLEW?
where x = 1 or 2 is the module number
(default - 1)
Short form: :SOUR:SLEW?
Function: Query rate value
Response: Decimal number representing rate value in selected units/second.
Description
When the controller rate is selected as value, this command is used to set the controllers
rate in selected units/second.
Example
TX> :SOUR:PRES:SLEW?
RX> :SOUR:PRES:SLEW 2.0000000
Change current unit of mbar to bar and query again:
TX> :SOUR:PRES:SLEW?
RX> :SOUR:PRES:SLEW 0.0020000
To set slew rate to other values, such as:
TX> :SOUR:PRES:SLEW 4
TX> :SOUR:PRES:SLEW?
RX> :SOUR:PRES:SLEW 4.0000000
TX> :SOUR:PRES:SLEW max
TX> :SOUR:PRES:SLEW?
RX> :SOUR:PRES:SLEW 99999999.0000000
TX> :SOUR:PRES:SLEW min
TX> :SOUR:PRES:SLEW?
RX> :SOUR:PRES:SLEW 0.0

Table of Contents

Other manuals for GE PACE5000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE PACE5000 and is the answer not in the manual?

GE PACE5000 Specifications

General IconGeneral
BrandGE
ModelPACE5000
CategoryTest Equipment
LanguageEnglish

Related product manuals