PACE Series SCPI Manual
K0472 Revision A 4-41
SENS:AIRF:QFE
Applicability: PACE1000
Command Syntax
n/a
Parameter:
Short form:
Function:
Query Syntax
SENSe:AIRField:QFE?
Short form: SENS:AIRF:QFE?
Function: This query reads the QFE value when the instrument is in airfield mode.
Response: A decimal reading in selected units of pressure.
Description
Queries the QFE reading, returns the QFE reading in selected units of pressure when the
instrument is in airfield mode.
Note: QFE is the mean sea level pressure corrected for temperature and adjusted for a specific
site (i.e. a datum such as an airfield).