4 Command and Query Summary
K0472 Revision A 4-8
:CAL:PRES:ZERO:AUTO
Applicability: PACE1000, PACE5000, PACE6000
Command Syntax
:CALibration[x]:[PRESsure]:ZERO:AUTO <Boolean>
where: x (module) = 1 or 2 (default - 1)
Parameter: <boolean>
0 - aborts a zero process
1 - starts a zero process
Short form: :CAL:ZERO:AUTO
Function: Pressure zeroing
Default: 0
Query Syntax
:CALibration[x]:[PRESsure]:ZERO:AUTO?
where: x (module) = 1 or 2 (default - 1)
Short form: :CAL:ZERO:AUTO?
Function: Query progress of zero
Response:
0 - Zero complete or not in progress.
1 - Zero in progress.
Description
This command starts or aborts a zero process. The progress of the zero can be monitored by
using the query.