PACE Series SCPI Manual
K0472 Revision A 4-83
:SOUR:MACH:LEV:IMM:AMPL
Applicability: PACE6000
Command Syntax
:SOURce[x]:MACH[:LEVel][:IMMediate][:AMPLitude] <number>
where x = 1 or 2 is the module number
(default - 1)
Parameter: <decimal> in MACH number.
Short form: SOURce:MACH <number>
Function: Sets the MACH set-point
Default: 0.0
Query Syntax
:SOURce[x]:MACH[:LEVel][:IMMediate][:AMPLitude] <number>?
Short form SOURce:MACH?
Function: Programmable set-point value
Response: Decimal number representing the set-point in MACH number.
Description
This command sets the MACH set-point. This query gets the decimal number representing
the set-point in MACH number.
Note: If the instrument is not in aero mode it returns an error: -221. Settings conflict.