4 Command and Query Summary
K0472 Revision A 4-86
:SOUR:PRES:INL
Applicability: PACE5000, PACE6000
Command Syntax
:SOURce[x][:PRESsure]:INLimits <number>
where x = 1 or 2 is the module number
(default - 1)
Parameter: <decimal> in limits value as % full-scale
Short form: :SOUR:INL <number>
Function: The controller has an in-limits set-point indicator. This can generate a
service request when the pressure is within limits for a set time period.
Default: 0.01 % full-scale
minimum 0% full-scale
maximum 100% full-scale
Query Syntax
:SOURce[x][:PRESsure]:INLimits?
where x = 1 or 2 is the module number
(default - 1)
Short form: :SOUR:INL?
Function: Query in-limits value
Response: Decimal representing in-limits value as % full-scale.
Description
Sets the in-limits value.
Example
TX> :SOUR:PRES:INL?
RX> :SOUR:PRES:INL 0.0200000
The current in-limits set-point is 0.02% full-scale
Can be set to:
TX> :SOUR:PRES:INL 0.01
TX> :SOUR:PRES:INL?
RX> :SOUR:PRES:INL 0.0100000