EasyManua.ls Logo

GE PACE5000

GE PACE5000
176 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PACE Series SCPI Manual
K0472 Revision A 4-123
:SYST:PASS:CEN:STAT
Applicability: PACE1000, PACE5000, PACE6000
Command Syntax
n/a
Parameter:
Short form:
Function:
Query Syntax
:SYSTem:PASSword:CENable:STATe?
Short form: SYST:PASS:CEN:STAT?
Function: Gets the status of the calibration password.
Response: 0 - disabled
1- enabled
Description
This query asks if the calibration is enabled or disabled.
Example
It can be enabled by sending command and password
TX> :SYST:PASS:CEN 2317100
TX> :SYST:PASS:CEN:STAT?
RX> :SYST:PASS:CEN:STAT 1
When enabled and a 3-point calibration can be carried out.
It can be disabled again by issuing a command and a password, such as:
TX> :SYST:PASS:CDIS 2317100
TX> :SYST:PASS:CEN:STAT?
RX> :SYST:PASS:CEN:STAT 0