EasyManuals Logo

GE PACE5000 Technical Manual

GE PACE5000
176 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #20 background imageLoading...
Page #20 background image
3 Status System
K0472 Revision A 3 - 2
The status system implemented in the instrument is shown in the following diagram:
Note:
Initial values of registers are 0, with the queues empty.
Key:
C = Condition - variable values
EV = Latched values
EN = Bit mask
Figure 3-1 Status System
logical OR
MAV
ESB
OSB
MSS
MAV
ESB
OSB
Status Byte
EN 8 bitEV
*SRE
STAT:OPER
:ENAB
*STB?
SERIAL POLL
STAT:OPER
:PRES:ENAB 511
Pressure Operations 15 bit
STAT
:OPER
:EVENT?
STAT:OPER
:PRES:ENAB?
Operation Status 16 bit
logical OR
CEV EN
logical OR
C
EVEN
Output Queue
Message
Message
Message
Message
SYST:ERR?
*SRE
Error Queue
Error Message
Error Message
Error Message
Error Message
Error Message
logical OR
QYE
CME
EXE
QYE
CME
EXE
Standard Event
EV EN 8 bit
*ESR? *ESE
MSS
EAV
EAV