EasyManua.ls Logo

GE PACE5000

GE PACE5000
176 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4 Command and Query Summary
K0472 Revision A 4-18
Description
This query returns a string representing the range, a number representing the upper full-
scale and lower full-scale.
Example
for a dual module instrument, both modules with the barometric option fitted.
TX> :INST:LIM?
RX> :INST:LIM "7.00barg", 7350.0000000, -1100.0000000
TX> :INST:LIM1?
RX> :INST:LIM "7.00barg", 7350.0000000, -1100.0000000
TX> :INST:LIM2?
RX> :INST:LIM2 "20.00barg", 21000.0000000, -1100.0000000
TX> :INST:LIM3?
RX> :INST:LIM3 "2.00barg", 2100.0000000, -1100.0000000
TX> :INST:LIM4?
RX> :INST:LIM4 "BAROMETER", 1207.5000000, 825.0000000
TX> :INST:LIM5?
RX> :INST:LIM5 "3.50barg", 3675.0000000, -1100.0000000
TX> :INST:LIM6?
RX> :INST:LIM6 "20.00barg", 21000.0000000, -1100.0000000
TX> :INST:LIM7?
RX> :INST:LIM7 "10.00barg", 10500.0000000, -1100.0000000
TX> :INST:LIM8?
RX> :INST:LIM8 "BAROMETER", 1365.0000000, 38.5000000

Table of Contents

Other manuals for GE PACE5000

Related product manuals