Chapter 5
Staff/duty stations
Telligence Patient-Staff Communication System Operation Manual 109
2. To cancel a call from this station (that was not already cleared
by a staff member answering the call), press the Cancel button
on the originating call station, or on any associated patient or
staff room station.
Call button lights extinguish.
Dome lights extinguish.
Tones stop sounding at staff consoles and annunciators,
and at stations where staff follow is active.
The nurse call row disappears from staff consoles and
annunciators.
If the nurse call had a reminder set for it, the following happens
when the Cancel button is pressed:
Dome light display of the staff level reminder ceases.
The reminder row disappears from staff console or
annunciator Reminders screens.
Remote cancel stations
The model HC-PB3-Cancel remote cancel station (see Figure 42 on
page 109) lets you cancel nurse calls from convenient locations in
patient or staff rooms, instead of trying to press a Cancel button at a
patient or staff/duty station that is blocked by medical equipment.
When a remote cancel station‘s Cancel button is linked with other
associated stations for a bed or a room, all calls placed from the bed
or room can be immediately canceled from this station.
Figure 42: Remote cancel station (model HC-PB3-Cancel, shown without
pull cord)
A remote cancel station is typically associated with a patient station,
or other peripheral stations. The station, which can be equipped
with a pull cord, may be installed in patient rooms or other rooms

Table of Contents