Chapter 5
Staff/duty stations
Telligence Patient-Staff Communication System Operation Manual 109
2. To cancel a call from this station (that was not already cleared
by a staff member answering the call), press the Cancel button
on the originating call station, or on any associated patient or
staff room station.
• Call button lights extinguish.
• Dome lights extinguish.
• Tones stop sounding at staff consoles and annunciators,
and at stations where staff follow is active.
• The nurse call row disappears from staff consoles and
annunciators.
If the nurse call had a reminder set for it, the following happens
when the Cancel button is pressed:
• Dome light display of the staff level reminder ceases.
• The reminder row disappears from staff console or
annunciator Reminders screens.
Remote cancel stations
The model HC-PB3-Cancel remote cancel station (see Figure 42 on
page 109) lets you cancel nurse calls from convenient locations in
patient or staff rooms, instead of trying to press a Cancel button at a
patient or staff/duty station that is blocked by medical equipment.
When a remote cancel station‘s Cancel button is linked with other
associated stations for a bed or a room, all calls placed from the bed
or room can be immediately canceled from this station.
Figure 42: Remote cancel station (model HC-PB3-Cancel, shown without
pull cord)
A remote cancel station is typically associated with a patient station,
or other peripheral stations. The station, which can be equipped
with a pull cord, may be installed in patient rooms or other rooms