EasyManua.ls Logo

GE Telligence Patient-Staff Communications System - Clearing Reminders

GE Telligence Patient-Staff Communications System
136 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 3
Using staff consoles and annunciators
Telligence Patient-Staff Communication System Operation Manual 39
1.
Select the Remind tab, and press the Connect button to answer
the first call listed, which is the call reminder with the highest
priority in the reminder list.
— or —
2. Select the Remind tab, touch the reminder row, and press the
Connect button to connect to the selected room or bed.
Clearing reminders
Reminders that have been set for normal priority calls can be
cleared from a staff console or annunciator by pressing the Clear
Reminder button on the Reminders screen.
Reminders for other call priorities can only be canceled by pushing
the Cancel button on the patient station where the call originated.
This clears the reminder from the nurse call system, so that it no
longer appears on staff consoles or annunciators, and Corridor
lights and zones lights stop flashing.
To clear a reminder on the staff console or annunciator:
1. Select the Remind tab.
2. Select the normal priority reminder row, and press the Clear
Reminder button.
Clear reminders only if you are certain the caller is no longer on the
line, and that a staff member has handled the original request.
Note: This does not cancel the call, but only clears the reminder.

Table of Contents

Related product manuals