Chapter 4
Patient stations
Telligence Patient-Staff Communication System Operation Manual 59
Staff normal calls
Assist light on the patient station
Call status lights on single and dual patient stations
Call assurance feedback tone at the patient station
Light sections in associated corridor and zone lights show a
steady white light.
Associated zone lights emit a very slow, repeating buzz.
Staff normal tones sound at staff consoles and annunciators,
and at stations where the staff follow function is active.
The call appears on the Call screen on staff consoles and
annunciators.
Patient calls
Patients place nurse calls via the patient station by:
Pressing the nurse call button on a Smart Pillow Speaker
Pressing a pushbutton or squeeze pad actuator on a call cord
Pressing the nurse call button on an electronic bed side-rail
control panel
Canceling a call from the patient station
Press the Cancel button on the patient station to cancel any
nurse call placed from the patient station.
— or —
Press any linked Cancel button to cancel calls from the patient
station.
Pressing Cancel on or for the patient station causes the
following to happen:
Dome lights and patient room lights extinguish.
Tones sounding at staff consoles/annunciators and staff follow-
activated stations cease.
The Call screen nurse call row and any corresponding
Reminders screen reminder row on staff consoles and
annunciators disappear.
Pressing Cancel cancels all calls in the room but lavatory calls.
Notes
Normal priority calls are canceled when answered at a staff
console or annunciator.