Chapter 4
Patient stations
84 Telligence Patient-Staff Communication System Operation Manual
pulling the pull cord, a patient or staff member places a lavatory
emergency call.
Pressing the Help button places a patient normal call for staff
assistance in the patient bathroom. This station is not capable of
lighting a dome light by the bathroom inside the patient room.
Figure 33: Two-priority lavatory station (shown with pull cord)
Placing a lavatory emergency call
1. Press the Emerg button or pull the station‘s cord to summon
emergency assistance from staff members. Doing so causes
this to happen:
• The station‘s Emerg light turns on.
• Any associated patient or staff/duty station‘s Call lights turn
on.
• The call assurance tone sounds at any associated patient
or staff/duty station.
• Top light sections of associated corridor and zone lights fast
flash red.
• The lavatory emergency tone (slow beeping) sounds at staff
consoles and all annunciators, at associated stations in the
patient room when presence is registered, and at stations
where staff follow is activated.
• A lavatory emergency nurse call row appears on the Call
screens on staff consoles and all annunciators.
2. Press the Cancel button on the original lavatory station, or on
any associated station in the patient bathroom, to cancel the
lavatory emergency call.