B
Brian FloresAug 5, 2025
Why are my Genrui Laboratory Equipment test results unstable?
- MMichelle GreenAug 5, 2025
Unstable test results in your Genrui Laboratory Equipment may occur due to several reasons: * Absorbance wasn't tested during startup – test absorbance during system maintenance. * The reagent is ineffective – exchange it for effective reagents. * The aspiration volume is insufficient – adjust the aspiration volume to be at least 600 μL. * The standard wasn't tested – measure the blank and standard again. * The wavelength is unavailable – use another wavelength to measure again. * Air is aspirated carelessly during operation – operate carefully according to the related requirements.