WP21 Series Operation Manual Chapter 1 Brief Introduction
2
Absorbance
stability
At 340nm, the absorbance change in 20
minutes should not be greater than 0.005.
Absorbance
repeatability
It is expressed with coefficient of variation
(CV) and the value should not be greater
than 1.0%
Temperature
accuracy and
fluctuation
Temperature deviation ≤ ±0.3,
fluctuation ≤ ±0.2
Sample
carryover rate
1%
Intra-batch
clinical
precision
(CV%)
ALT ≤ 5%, TP ≤ 2.5%, UREA ≤ 3.5%
Basic parameters
Wavelength selector
Filtering
Type of absorption cells
Flow cells
Analysis methods
End-point, two-point, kinetic
methods, etc.
Wavelength
340, 380, 405, 505, 546, 578, 620
and 700nm (optional)
Power
180VA
Net weight
About 6kg