APPENDIX
207
Model
ASR-2050
ASR-2050R
ASR-2100
ASR-2100R
AC mode output ratings (AC rms)
Voltage
Setting
Range
*1
0.0 V to 175.0 V / 0.0 V to 350.0 V
Setting
Resolution
0.1 V
Accuracy
*2
±(0.5 % of set + 0.6 V / 1.2 V)
Output phase
Single phase, Two-wire
Maximum
current
*3
100 V
5 A
10 A
200 V
2.5 A
5 A
Maximum
peak current
*4
100 V
20 A
40 A
200 V
10 A
20 A
Load power factor
0 to 1 (leading phase or lagging phase)
Power capacity
500 VA
1000 VA
Frequency
Setting range
AC Mode: 40.00 Hz to 999.9 Hz, AC+DC Mode:
1.00 Hz to 999.9 Hz
Setting
resolution
0.01 Hz (1.00 to 99.99 Hz), 0.1 Hz (100.0 to 999.9
Hz)
Accuracy
For 45 Hz to 65 Hz: 0.01% of set
For 1 Hz to 999.9 Hz: 0.02% of set
Stability
*5
± 0.005%
Output on/off phase
0.0° to 359.9° variable (setting resolution 0.1°)
DC offset
*6
Within ± 20 mV (TYP)
*1
100 V / 200 V range
*2
For an output voltage of 17.5 V to 175 V / 35 V to 350 V, sine wave, an output
frequency of 45 Hz to 65 Hz, no load, DC voltage setting 0V (AC+DC mode)
and 23°C ± 5°C
*3
For an output voltage of 1 V to 100 V / 2 V to 200 V. Limited by the power
capacity when the output voltage is 100 V to 175 V / 200 V to 350 V. If there
is the DC superimposition, the current of AC+DC mode satisfies the
maximum current. In the case of lower than 40 Hz, and the ambient
temperature is 40°C or higher, the maximum current will be decrease.
*4
With respect to the capacitor-input rectifying load. Limited by the maximum
current.
*5
For 45 Hz to 65 Hz, the rated output voltage, no load and the resistance load
for the maximum current, and the operating temperature.
*6
In the case of the AC mode and output voltage setting to 0 V.