GPT-9000/9000A Series User Manual
140
Example
MANU:DCW:TTIM 1
Sets the DCW test time to 1 second.
MANU:DCW:REF
Set
Query
Description
Sets or returns the DCW reference value in mA.
The test must first be in DCW mode before this
command can be used.
The reference value must be less than the HI SET
value.
The reference value uses the same range as the HI
SET value.
Syntax
Query Syntax
MANU:DCW:REF <NR2>
MANU:DCW:REF?
Parameter/
Return parameter
<NR2>
0.000 ~ 010.9 (GPT-98XX)
0.000 ~ 020.9 (GPT-99XX/99XXA)
Example
MANU:DCW:REF 0.01
Sets the DCW reference to 0.01 mA.
MANU:DCW:ARCCurrent
Set
Query
Description
Sets or returns the DCW ARC current value in mA.
ARC must be enabled to set the ARC current. The
test must first be in DCW mode before this
command can be used.
ARC current uses the same range as the HI SET
value. The ARC current is limited to 2X the HI SET
value.
Syntax
Query Syntax
MANU:DCW:ARCCurrent <NR2>
MANU:DCW:ARCCurrent?
Parameter/
Return parameter
<NR2>
1.000 ~ 20.00 (GPT-98XX)
2.000 ~ 040.0 (GPT-99XX/99XXA)
w ww . . co m
information@itm.com1.800.561.8187
All manuals and user guides at all-guides.com

Table of Contents

Other manuals for GW Instek GPT-9000 series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GW Instek GPT-9000 series and is the answer not in the manual?

GW Instek GPT-9000 series Specifications

General IconGeneral
BrandGW Instek
ModelGPT-9000 series
CategoryTest Equipment
LanguageEnglish

Related product manuals