APPENDIX
171
Ground Bond Test
Output Current Range
03.00A~30.00A (GPT-98XX)
03.00A~32.00A (GPT-99XX)
Output Current Accuracy
(1% of setting +0.2A) when 3AI8A
(1% of setting +0.05A) when 8A<I30A
(GPT-98XX)
(1% of setting +0.05A) when 8A<I32A
(GPT-99XX)
Output Current Resolution
0.01A
Frequency
50Hz/60Hz selectable
Ohmmeter Measurement
Accuracy
(1% of reading +2m)
Ohmmeter Measurement
Range
10m~650.0m (depending on output current)
GPT-98XX:
GPT-99XX:
Test Voltage
Max. 6V(AC)open-circuit
Ohmmeter Measurement
Resolution
0.1m