EasyManua.ls Logo

GW Instek GPT-9000 series - Page 171

GW Instek GPT-9000 series
182 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
APPENDIX
171
Ground Bond Test
Output Current Range
03.00A~30.00A (GPT-98XX)
03.00A~32.00A (GPT-99XX)
Output Current Accuracy
(1% of setting +0.2A) when 3AI8A
(1% of setting +0.05A) when 8A<I30A
(GPT-98XX)
(1% of setting +0.05A) when 8A<I32A
(GPT-99XX)
Output Current Resolution
0.01A
Frequency
50Hz/60Hz selectable
Ohmmeter Measurement
Accuracy
(1% of reading +2m)
Ohmmeter Measurement
Range
10m~650.0m (depending on output current)
GPT-98XX:
GPT-99XX:
Test Voltage
Max. 6V(AC)open-circuit
Ohmmeter Measurement
Resolution
0.1m
Continued…
w ww . . co m
information@itm.com1.800.561.8187
All manuals and user guides at all-guides.com
all-guides.com

Table of Contents

Other manuals for GW Instek GPT-9000 series

Related product manuals