5. The trimming starts and ends automatically.
SHORT CIRCUIT TRIM
IN PROGRESS
ALL FREQUENCY
Trim pass
The display goes back to the
calibration mode menu.
Trim fail
The Fail message appears on the
display. Press any other key to go
back to the original menu.
Press any key to continue...
FAILED
SHORT CIRCUIT TRIM
6. Switch the fixture setting from O/C to S/C (or
from S/C to O/C) and repeat from step 1.
Note
When an “O/S Trim Error!” message appears, it
means either that you haven’t properly performed an
O/S trimming yet or that there is not an exactly
matched calibration factor that can be used to
calculate the most precise measurement reading
based on the spot frequency you are using. When an
“O/S Trimming Error!” message appears, the LCR-
8000G will use an existing calibration factor at the
closest frequency point to the spot frequency you are
w ww . . co m