CALIBRATION
129
using to calculate a measurement. However, this
measurement is only an approximation. To obtain the
most optimized measurement accuracy and to clear
the warning message, it is best to perform O/S
trimming again, based on the actual measuring
frequency being used.
To get the most accurate measurement reading for the
frequency spot you are currently using, any of the O/S
trimming options (All-frequency, 10kHz, 100kHz
and Spot frequency) can be used despite the differing
number of total calibrated frequency points, see O/S
Trimming Points section in the Appendix for details on
page 132.