FUNCTION MENU
155
Results:
Power Source
OPP tripped
Exit
OPPUSB
01/Oct/2018
OPP Test
4.81
V
10.530
A
W
50.61
Fetch Watt: Fetch Volt:
60.76 4.79
50.71 4.82
40.57 4.83
30.40 4.85
Watt:
50.72
0:00:07
50.71 W
Save
TEST
Result
The OPP Test will return the current setting of the
last step when the power supply’s OPP was
tripped.
Power Source
OPP time out
Exit
OPPUSB
01/Oct/2018
OPP Test
4.81
V
10.525
A
W
50.61
Fetch Watt: Fetch Volt:
91.21 4.76
81.01 4.76
70.91 4.77
60.73 4.79
Time Out
50.61
0:00:15
OPP time out will occur if the power supply’s OPP
fails to trigger. This is determined when the
measured voltage is less than Trig V and the
measured watt is greater than End W.