PSW Series User Manual
32
Abort Commands
A
BORt ....................................................................................32
ABORt
Set
Description
The ABORt command will cancel any triggered
actions.
Syntax ABORt
APPLy Commands
A
PPL
y
.....................................................................................32
APPLy
Set
Query
Description
The APPLy command is used to set both the
voltage and current. The voltage and current will
be output as soon as the function is executed if the
programmed values are within the accepted range.
An execution error will occur if the programmed
values are not within accepted ranges.
The Apply command will set the voltage/current
values but these values will not be reflected on the
display until the Output is On or if the
DISPlay:MENU:NAME 3 (set menu) command is
used.
Syntax APPLy {<voltage>|MIN|MAX}[,{<current>|MIN|MAX}
APPLy?
Parameter <voltage>
<NRf> 0% ~ 105% of the rated output
voltage.
<current>
<NRf> 0% ~ 105% of the rated output
current.
MIN
0 volts/0 amps