REMOTE CONTROL
37
OUTPut:DELay:OFF
Set
Query
Description
Sets the Delay Time in seconds for turning the
output off. The delay is set to 0.000 by default.
Syntax
Return Syntax
OUTPut:DELay:OFF <NRf>
OUTPut:DELay:OFF?
Parameter <NRf>
0.00~99.99 seconds, where 0=no delay.
Return parameter <NRf>
Returns the delay off time in seconds until the
output is turned off.
OUTPut:MODE
Set
Query
Description
Sets the PSW output mode. This is the equivalent
to the F-03 (V-I Mode Slew Rate Select) settings.
Syntax
Return Syntax
OUTPut:MODE {<NR1>|CVHS|CCHS|CVLS|CCLS}
OUTPut:MODE?
Parameter 0
CV high speed priority
CVHS
CV high speed priority
1
CC high speed priority
CCHS
CC high speed priority
2
CV slew rate priority
CVLS
CV slew rate priority
3
CC slew rate priority
CCLS
CC slew rate priority
Return parameter <NR1>
Returns the output mode.
OUTPut[:STATe][:IMMediate]
Set
Query
Description
Turns the output on or off.
Syntax
Query Syntax
OUTPut[:STATe][:IMMediate] { OFF | ON | 0 | 1 }
OUTPut[:STATe][:IMMediate]?
Parameter 0
<NR1> Turns the output off.
OFF
Turns the output off.
1
<NR1> Turns the output on.