[SOURce:]CURRent[:LEVel]:TRIGgered
[:AMPLitude]
Set
Query
Description
Sets or queries the current level in amps when a
software trigger has been generated.
Syntax
[SOURce:]CURRent[:LEVel]:TRIGgered[:AMPLitude]
{<NRf>|MIN|MAX}
Query Syntax [SOURce:]CURRent[:LEVel]:TRIGgered[:AMPLitude]?
{<NRf>|MIN|MAX}
Parameter/Return
<NRf>
0%~105% of the rated current output in amps.
MIN
Minimum current level.
MAX
Maximum current level.
Example SOUR:CURR:LEV:TRIG:AMPL? MAX
37.800
Returns the maximum possible current level in amps.
[SOURce:]CURRent:PROTection[:LEVel]
Set
Query
Description
Sets or queries the OCP (over-current protection)
level in amps.
Syntax
[SOURce:]CURRent:PROTection[:LEVel]
{<NRf>|MIN|MAX}
Query Syntax [SOURce:]CURRent:PROTection[:LEVel]? {MIN|MAX}
Parameter/Return
<NRf>
10%~110% of the rated current output level.
MIN
Minimum current level.
MAX
Maximum current level.
Example SOUR:CURR:PROT:LEV? MIN
+3.600
Returns the minimum possible current level in amps.

Table of Contents

Other manuals for GW Instek PSW Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GW Instek PSW Series and is the answer not in the manual?

GW Instek PSW Series Specifications

General IconGeneral
Ripple & Noise (CC)≤ 3mArms
ProtectionsOVP, OCP, OTP
Remote SenseYes
ProgrammableYes
Operating Temperature0°C to 40°C
Storage Temperature-20°C to 70°C
Voltage Resolution1mV
Current Resolution1mA
Current Accuracy±0.1% + 10mA
Ripple & Noise (CV)≤1mVrms
Load Regulation (CC)≤0.1% + 5mA
Line Regulation (CC)≤0.1% + 5mA
Remote InterfaceUSB, RS232