REMOTE CONTROL
43
Source Commands
[SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude] 43
[SOURce:]CURRent[:LEVel]:TRIGgered[:AMPLitude] .. 44
[SOURce:]CURRent:PROTection[:LEVel] ....................... 44
[SOURce:]CURRent:SLEW:RISing .................................... 45
[SOURce:]CURRent:SLEW:FALLing ................................ 45
[SOURce:]RESistance[:LEVel][:IMMediate]
[:AMPLitude] .......................................................................... 46
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude] 46
[SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude] . 47
[SOURce:]VOLTage:PROTection[:LEVel] ....................... 47
[SOURce:]VOLTage:SLEW:RISing .................................... 47
[SOURce:]VOLTage:SLEW:FALLing ............................... 48
[SOURce:]CURRent[:LEVel][:IMMediate]
[:AMPLitude]
Set
Query
Description
Sets or queries the current level in amps. For
externally set current levels (from the analog
control connector) the set current level is returned.
Syntax [SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude]
{<NRf>|MIN|MAX}
Query Syntax [SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude]?
{MIN|MAX}
Parameter/Return
<NRf>
0~105% of the rated current output level.
MIN
Minimum current level.
MAX
Maximum current level.
Example SOUR:CURR:LEV:IMM:AMPL? MAX
37.800
Returns the maximum possible current level in amps.