REMOTE CONTROL
45
[SOURce:]CURRent:SLEW:RISing
Set
Query
Description
Sets or queries the rising current slew rate. This is
only applicable for CC slew rate priority mode.
Syntax
Query Syntax
[SOURce:]CURRent:SLEW:RISing {<NRf>|MIN|MAX}
[SOURce:]CURRent:SLEW:RISing? {MIN|MAX}
Parameter/Return
<NRf>
0.01A/s~200% (PSW 30-36)
0.1A/s~200% (PSW 30-72)
0.1A/s~200% (PSW 30-108)
0.01A/s~200% (PSW 80-13.5)
0.01A/s~200% (PSW 80-27)
0.01A/s~200% (PSW 80-40.5)
MIN
Minimum rising current slew rate.
MAX
Maximum rising current slew rate.
Example SOUR:CURR:SLEW:RIS 72
Sets the rising current slew rate to 72A/s.
[SOURce:]CURRent:SLEW:FALLing
Set
Query
Description
Sets or queries the falling current slew rate. This is
only applicable for CC slew rate priority mode.
Syntax [SOURce:]CURRent:SLEW:FALLing {<NRf>|MIN|MAX}
Query Syntax [SOURce:]CURRent:SLEW:FALLing? {MIN|MAX}
Parameter/Return
NRf
0.01A/s~200% (PSW 30-36)
0.1A/s~200% (PSW 30-72)
0.1A/s~200% (PSW 30-108)
0.01A/s~200% (PSW 80-13.5)
0.01A/s~200% (PSW 80-27)
0.01A/s~200% (PSW 80-40.5)
MIN
Minimum falling current slew rate
MAX
Maximum falling current slew rate
Example SOUR:CURR:SLEW:FALL 1
Sets the falling current slew rate to 1A/s.