OXIDATION PROGRAM 2 TC
BASE IN 1 12p
BASE IN 1 specifies the quantity of base reagent (in pulses) to
be injected into the reactor for the Pre Oxidation phase in TC
mode at range 2.
SAMPLE IN 5p , M-V
SAMPLE IN defines the amount of sample (in pulses) to be
injected into the reactor in TC mode at range 2.
BASE IN 2 3p
BASE IN 2 identify the quantity of base reagent (in pulses) to be
injected into the reactor for the VOC Oxidation phase in TC
mode at range 2.
TOC ACID 15p
TOC ACID states the amount of acid (in pulses) to be injected
into the reactor for the TOC SPARGE phase in TC mode at
range 2.
OXIDATION PROGRAM 2 VOC
All system parameters for OXIDATION PROGRAM 2 VOC
analysis mode are covered in Oxidation Program 2 TC and
Oxidation Program 2 TIC + TOC above.
OXIDATION PROGRAM 2 TC - TIC
TC SAMPLE IN 5p , M-V
TC SAMPLE IN specifies the amount of sample (in pulses) to be
injected into the TC reactor at range 2 in TC - TIC systems.
TIC SAMPLE IN 5p , P-V
TIC SAMPLE IN specifies the amount of sample (in pulses) to be
injected into the TIC reactor at range 2 in TC - TIC systems.
TIC ACID 3p
TIC ACID determines the amount of acid reagent (in pulses) to
be injected into the TIC reactor at range 2 in TC - TIC systems.
TOC ACID 12p
TOC ACID states the amount of acid reagent (in pulses) to be
injected into the TC reactor at range 2 in TC - TIC systems.
OXIDATION PROGRAM 3
OXIDATION PROGRAM 3 TIC + TOC
RANGE CHANGE 2-3 150.0 mgC/l
TN 75.0mgN/l
RANGE CHANGE 2-3 defines the concentration level for the
BioTector to change its range automatically from range 2 to
range 3.
The RANGE CHANGE 2-3 is typically the top concentration
point of TOC/TN RANGE 2 in System Range Data menu by
default.
RANGE CHANGE 3-2 120.0mgC/l
TN 60.0mgN/l
RANGE CHANGE 3-2 specifies the concentration level for the
BioTector to change its range automatically from range 3 to
range 2.
The RANGE CHANGE 3-2 is typically 20% less of the top
concentration point of TOC/TN RANGE 2 in System Range Data
menu by default.
RANGE CHANGE 3-1 40.0mgC/l
TN 20.0mgN/l
RANGE CHANGE 3-1 defines the concentration level for the
BioTector to change its range automatically from range 3 to
range 1.
The RANGE CHANGE 3-1 is typically 20% less of the top
concentration point of TOC/TN RANGE 1 in System Range Data
menu by default.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach BioTector B7000 and is the answer not in the manual?

Hach BioTector B7000 Specifications

General IconGeneral
BrandHach
ModelBioTector B7000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals