SAMPLE IN states the amount of sample (in pulses) to be
injected into the reactor at range 3. If the setting is 0p MF-180, it
means that the sample injection at this range will be carried out
by means of the Motorized Fixed 180° rotation of the Sample
(ARS) Valve.
When SAMPLE IN setting is MF-180, the Sample Pump does
not inject the sample into reactor. Instead acid reagent (or base
reagent depending on the analysis type of the system) and the
carrier gas flow is used to inject the fixed volume of the sample
sitting inside the Sample (ARS) Valve.
TIC ACID defines the amount of acid reagent (in pulses) to be
injected into the reactor at range 3.
TOTAL BASE IN specifies the amount of base (in pulses) to be
injected into the reactor for the BASE OXIDATION phase at
range 3.
If available, BASE OXIDATION defines the oxidation time (150s
by default), which can be programmed independently for range
3.
If the Base Oxidation time is different than the one programmed
in Oxidation Program 1 menu, depending on the difference
between the two settings, the reaction time at range 3 will be
either longer or shorter than the reaction time displayed in
Oxidation Program 1 menu.
TOC ACID states the amount of acid (in pulses) to be injected
into the reactor for the TOC SPARGE phase at range 3.
BASE IN 1 specifies the quantity of base reagent (in pulses) to
be injected into the reactor for the Pre Oxidation phase in TC
mode at range 3.
SAMPLE IN defines the amount of sample (in pulses) to be
injected into the reactor in TC mode at range 3.
BASE IN 2 identifies the quantity of base reagent (in pulses) to
be injected into the reactor for the VOC Oxidation phase in TC
mode at range 3.
TOC ACID states the amount of acid (in pulses) to be injected
into the reactor for the TOC SPARGE phase in TC mode at
range 3.
All system parameters for OXIDATION PROGRAM 3 VOC
analysis mode are covered in Oxidation Program 3 TC and
Oxidation Program 3 TIC + TOC above.
OXIDATION PROGRAM 3 TC - TIC
TC SAMPLE IN specifies the amount of sample (in pulses) to be
injected into the TC reactor at range 3 in TC - TIC systems.
TIC SAMPLE IN specifies the amount of sample (in pulses) to be
injected into the TIC reactor at range 3 in TC - TIC systems.
TIC ACID determines the amount of acid reagent (in pulses) to
be injected into the TIC reactor at range 3 in TC - TIC systems.
TOC ACID states the amount of acid reagent (in pulses) to be
injected into the TC reactor at range 3 in TC - TIC systems.