Replace the 3.5mm OD, 1.5mm ID EMPP 562 tube in the HCl Water Pump if
installed in systems built with FMI Heavy Duty Circulation Pump (see item A in
service kit). If there are no HCl Water In and HCl Water Out ports on the
BioTector, this item does not apply.
_____
_____
Check that the peristaltic pumps are pumping correctly. The pump rate measurements are used to
confirm that the correct tubing has been installed in the appropriate pump. Any variation between
tube batches is corrected when the zero and span calibration is carried out.
Sample Pump rate for WMM60 Pump at 16 pulses should be between 5.5ml
and 7.5ml in ~8 seconds.
_____
_____
Acid Pump rate for SR25 Pump at 20 pulses should be between 3.9ml and
4.9ml in ~13 seconds.
_____
_____
Base Pump rate for SR25 Pump at 20 pulses should be between 3.9ml and
4.9ml in ~13 seconds.
_____
_____
Important Note: For the correct operation of the system, the measured Acid and Base Pump rates
must be identical or as close as possible. The maximum allowable difference in the measured
volumes for acid and base injections above is typically 5% or ~0.2ml. This figure is based on
standard system operation conditions and may not cover systems built with specific configurations.
HCl Water Pump (if installed) rate should be between 8.5ml and 11.5ml in ~18
seconds. If there are no HCl Water In and HCl Water Out ports on the
BioTector, this item does not apply.
_____
_____
Check that the valves are operating correctly:
Sample (ARS) Valve checked. MV4
_____
_____
Sample Out Valve checked. MV5
_____
_____
Purge Valve (if fitted) checked. MV51
In systems built with Multi Component Reactor (MCR), when Sample Out
Valve (MV5) is activated, the Purge Valve (MV51) automatically activates.
_____
_____
Exhaust Valve checked. MV1
_____
_____
TOC Acid Valve checked. MV6
_____
_____
Base Valve checked (if fitted). MV10
_____
_____
Injection Valve checked. MV7
_____
_____
Check all the other valves installed in the BioTector.
_____
_____
Check and confirm that MANUAL/CALIBRATION ports are not blocked or
isolated.
_____
_____
Remove Ozone Line Filter and wash it with DIW (or tap water). Dry the filter
well and reinstall in place.
_____
_____
Replace the 149mm filter mats in fan and vent housing (see item S in service
kit).
_____
_____
Check the operation of the fan.
_____
_____