7. Replace the indicator bottle as follows:
a. Remove the cap and seal from the indicator bottle and the brown DPD bottle.
b. Fill the brown DPD bottle approximately ¼ full with indicator solution.
c. Swirl the DPD bottle to mix.
d. Put the contents of the DPD bottle into the indicator bottle.
e. Invert the indicator bottle until all of the powder is dissolved (2 minutes).
f. Remove the used indicator bottle from the analyzer.
g. Put the indicator tubing in the new indicator bottle on the left side of the analyzer. Tighten the
cap.
8. Push enter.
9. When "TASK COMPLETE" shows on the display, push enter. The analyzer starts a
measurement cycle in approximately 30 seconds.
7.4 Clean the screen in the Y-strainer
Clean the screen in the Y-strainer when there is a blockage, which is identified with a LOW SAMPLE
FLOW warning. Complete the illustrated steps that follow.
22 English

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach Ultra Low Range CL17sc and is the answer not in the manual?

Hach Ultra Low Range CL17sc Specifications

General IconGeneral
Measurement MethodColorimetric DPD method
Power Requirements100-240 VAC, 50/60 Hz
Resolution0.01 mg/L
Accuracy±0.03 mg/L or ±5% of reading, whichever is greater
Sample Temperature5 to 50 °C
Cycle Time5 minutes
Sample Flow Rate200 mL/min
Outputs4-20 mA
Operating Temperature5 to 40 °C
Response Time5 minutes