E-25
Configuring the Spider8 with the Setup program Ý Channel-specific menus
Spider8
Explanation:
Signal edges:
If only 1 signal is measured, one signal edge is counted.
Signal edges and direction:
If 2 signals are measured: the first channel measures pulses (one
signal edge), the second channel defines the direction (ascending
or descending).
2-phase, single interpretation:
If two signals are measured, one edge of one signal is interpreted.
The phase position defines the direction.
2-phase, 4-fold interpretation
If two signals are measured, both edges of each signal are
interpreted. The phase position defines the direction.