106 HVAD® Instructions for Use
5.7 HeartWare
®
Monitor Care (continued)
Periodically or as needed:
The HeartWare
®
Monitor may be cleaned with:
Alcohol (Isopropyl 90% or Ethyl 70%)
 Silver, ethanol, acetic acid, peroxyacetic acid, hydrogen peroxide solution (active ingredients
of some sporicidal disinfecting wipes)
 UV-C disinfecting wand, one that radiates “short wave” UV-C band rays (100 to 280
nanometers) from a 4 watt bulb (or stronger).
The Monitor AC Adapter may be cleaned with the following agents:
Alcohol (Isopropyl 90% or Ethyl 70%)
Hydrogen peroxide solution (1.4%)
 n-Alkyl Dimethyl Dibenzyl Ammonium Chloride combined with n-Alkyl Dimethyl Ethybenzyl
Ammonium Chloride (active agent in some disinfecting wipes)
 Silver, ethanol, acetic acid, peroxyacetic acid, hydrogen peroxide solution (active ingredients
of some sporicidal disinfecting wipes)
 UV-C disinfecting wand, one that radiates “short wave” UV-C band rays (100 to 280
nanometers) from a 4 watt bulb (or stronger).
The Data Cable may be cleaned with the following agents:
Alcohol (Isopropyl 90% or Ethyl 70%),
Hydrogen peroxide solution (1.4%),
 n-Alkyl Dimethyl Dibenzyl Ammonium Chloride combined with n-Alkyl Dimethyl Ethybenzyl
Ammonium Chloride (active agent in some disinfecting wipes), or bleach solution (10%).
 Silver, ethanol, acetic acid, peroxyacetic acid, hydrogen peroxide solution (active ingredients
of some sporicidal disinfecting wipes)
 UV-C disinfecting wand,one that radiates “short wave” UV-C band rays (100 to 280
nanometers) from a 4 watt bulb (or stronger).
Product Disposal
HeartWare
®
Monitor
The HeartWare
®
Monitor contains a lithium battery (replaceable). Dispose of/recycle the
monitor’s internal battery in compliance with all applicable local, state, and federal laws and

WARNING! 
use alcohol or detergent on the monitor display. Gently wipe the display with a soft, lint free cloth.