124 HVAD® Instructions for Use
7.1 Postoperative Management (continued)
7.1.1 Setting Speed with HVAD
®
Pump
The pump speed for each patient should be individualized based on body surface area and
clinical condition.




 Speed should be maintained at a minimum level to: attain satisfactory hemodynamics (MAP

2
, end organ perfusion) and optimal decompression of
the heart without leftward shift of the intraventricular septum or suction.
 Echocardiography can be helpful when setting speed. It can provide detailed information

positioning.
One of the operating goals for the HVAD
®


®


waveform on the HeartWare
®


(see
gure below). Pulsatility is affected by a number of patient conditions including left ventricular
contractility, right heart function and left ventricular afterload.

®



For more information on HVAD
®
Pump operational guidelines, see Section 3.3.
NOTE: Recommended HVAD
®
Pump speeds are between 2400 RPM and 3200 RPM. HVAD
®
Pump speeds
outside this range may result in less than optimal HVAD
®
Pump operation.
Figure 111: HVAD
®

Figure 112: HeartWare
®
Monitor screen
showing HVAD
®
