54 HVAD® Instructions for Use
2.4 HVAD
®
Controller Power Sources (continued)
2.5 HeartWare
®
Battery Charger
The battery charger (Figure 32) is used to simultaneously recharge up to four batteries. It takes
approximately 5 to 6 hours to fully charge a depleted battery. Each battery slides into a bay and
is connected to the battery charger. It is safe to leave the batteries in the charger.
Figure 32
2.6 Carrying Cases and Shower Bag
The HeartWare
®
Waist Pack, HeartWare
®
Shoulder Pack, and HeartWare
®
Convertible Patient
Pack are used to safely secure, store and carry the controller and batteries. They can be used in

into each of the carrying cases.
For additional information on HeartWare
®
Batteries and AC/DC adapter, see Section 4.5.
For additional information on the battery charger, see Section 4.4.
CAUTION: The HeartWare
®
Waist Pack and the HeartWare
®
Shoulder Pack contain magnetic closures.

chest, including when sleeping. Per pacemaker and ICD manufacturer guidelines, magnets should
be kept at least 6 inches (15 centimeters) away from the pacemaker or ICD (please refer to
manufacturer guidelines for additional information).
WARNING! 
since this will stop the pump. At least one power source must be connected at all times.