59HeartWare™ HVAD™ System Pump
10
9
8
7
6
5
4
3
2
1
Appendix
Reference
Guides
Alarms and
Emergencies
Patient
Management
Surgical
Implant and
Explant
Monitor
Peripherals
and
Accessories
HVAD
®
PumpOverviewIntroduction
NOTE: Update hematocrit settings on the monitor whenever hematocrit changes by plus or minus 5%
or more.
Monitor and Controller Display Estimated Flow Range
“--“ Invalid, not available
“< -2.0 L/min“ less than -2.0 L/min
“-2.0 L/min” up to “10.0 L/min” -2.0 to 10.0 L/min
 greater than 10.0 L/min
NOTE: 
to variability of patient’s hematocrit.
3.2 Physiologic Control Algorithms
The HVAD
®
Pump control algorithms provide clinicians information about device performance
and HVAD
®

3.2.1 Flow Estimation
Estimated HVAD
®

hematocrit, based on a blood sample from the patient. The default hematocrit setting is 30%,

Adjustments to the hematocrit setting on the monitor should be made for hematocrit changes
of ± 5% or greater.

Controller Display messages and what they mean.
Table 22: Monitor and Controller Display Messages

Flow estimation accuracy can be maintained only if accurate hematocrit values are entered.



to thrombus or other materials (e.g. tissue fragments) in the device, due to an incorrect

WARNING! 
on the monitor or Controller Display of less than 2.0 L/min, or greater than 10.0 L/min may indicate an
electrical fault, incorrect hematocrit entry or an occlusion and/or thrombus or other materials (e.g. tissue
fragments) in the device. Inaccurate assessment of HVAD
®

treatment.