8.3 Medium Priority Alarms (continued)
The Following are Medium-Priority Alarms (continued):
[Low Flow]:
threshold.
For additional information about the Alarm Settings Tab, see Section 5.3.
[Suction]: The [Ventricular Suction Detection] alarm is triggered if the suction algorithm has

For additional information about the suction algorithm, see Section 3.2.2.
[Controller Fault]: The controller contains two microprocessors - one which controls pump
function (PMC) and a second which controls user interface functions (UIC) such as Controller
Display and buttons. The [Controller Fault] alarm indicates a possible controller malfunction may
have occurred, but during this fault the UIC processor still receives a heartbeat message from the
PMC indicating the PMC is still functioning and controlling the pump. The [Controller Fault] Alarm
will result in the word [Call] in the Controller Display, notifying the patient to call the clinician. The
clinician should query the patient about the frequency and duration of alarm as well as any

about any clinical symptoms/changes including dizziness, shortness of breath, angina and/or
palpitations. Based on the patient’s responses, the following course of action should be taken:
 If there was a single, isolated [Controller Fault] alarm with no change in pump or clinical
parameters, instruct the patient to report any additional alarms that may occur. Download

 Instruct the patient to return to the center as soon as reasonable (not emergently) so the

following situations occurs:
If a [Controller Fault] alarm has occurred and been resolved multiple times over a 24 hour
period, or
If a [Controller Fault] alarm has occurred in conjunction with other alarms even though it has
