Appendix B: Product Specifications (continued)
Software Parameters, Ranges & Factory Default Settings
Parameter Range Resolution Factory Default
Pump Speed 1,800 to 4,000 RPM 20 RPM 2,500 RPM
[Low Flow] alarm 1.0 to 9.9 L/min 0.1 L/min 1.0 L/min
[High Power] alarm
Limit
1.0 to 25.0 Watts 0.5 Watts 16.0 Watts
Suction Detection Off, Alarm  Off
Hematocrit 20-50% HCT 1% 30%
Data Logging 15 Minutes  15 Minutes
NOTE: All dimensions are given as length x width x height. PEEK is a registered trademark of Victrex plc.
Monitor REF1510 REF1520
Type 
Operating System  
Weight 2.5 kg 3.0 kg
Dimensions 29.9 x 23.5 x 4.5 cm
28.5 x 21.0 x 4.1 cm (without case)
29.9 x 29.9 x 6.4 cm (with case)
Electrical Ratings 19 V, 3 A maximum input 19 V, 3.4 A maximum input
Monitor AC Adapter
Weight 0.7 kg
Dimensions 16.6 x 9.6 x 5.6 cm
Electrical Ratings 100-240 V, 50-60 Hz, 215 VA input; 19 V, 4.7 A output
IEC 60601-1 Classifications
Type of protection against electric shock:





Degree of protection against electric shock:

Degree of protection against the ingress of water:






Recommended environmental conditions
Component Temperature Range Relative Humidity Atmospheric Pressure
Controller


Operating: 15% - 95% Operating: 700 -1060 hPa
Storage and Transport:

Storage and Transport:
10% - 93%
Storage and Transport:
500 -1060 hPa
Controller AC and
DC Adapter


Operating: 15% - 95% Operating: 700 -1060 hPa
Storage and Transport:

Storage and Transport:
10% - 93%
Storage and Transport:
500 -1060 hPa
Battery C
F)
Operating: 15% - 95% Operating: 700 -1060 hPa
Storage and Transport:

Storage and Transport:
10% - 93%
Storage and Transport:
500 -1060 hPa