Page 12 194412(11) - Hill-Rom® 900 Accella™ Bed - Instructions For Use
Electrical characteristics
Features
120V* 230V*
Voltage 120V AC 230V AC
Frequency: 60 Hz 50/60 Hz
Maximum absorbed power 500 VA 500 VA
Electric shock protection
Class I
Class according to IEC 60601-1
Type B
Protection against harmful ingress of water (according
t
o IEC 60529)
IPX4
Duty cycle 10% (2 min/18 min)
a
a. Do not operate electrical functions continuously for more than 2 minutes in any 18 minute period
when the bed is loaded at the safe working load value as this may damage electrical components.
The power supply of the actuator is temporarily cut off if the load factor is exceeded when using the
HiLow.
Chassis clearance (double-band 125
b
diameter casters) (C) 150 mm
a
Chassis clearance (double band 150
b
diameter casters) (C) 195 mm
a
Chassis clearance (150
b
diameter casters) (C) 203 mm
a
Head section
c
incline + 65°
a
Thigh section
c
incline + 28°
a
Foot section
c
incline - 3° to -22°
a
Trendelenburg/Reverse Trendelenburg
+ 17°/- 17°
a
Boost™ patient position mechanism
c
- 7 °/30 sec
Electric CPR - return adjustable head section (T1)
and sleep surface (T2) to flat position
T1 < 5 s
T2 < 30 s
Safe working load (SWL) 250 kg
Maximum patient weight
185-215 kg
d
LI900B4 bed weight (no mattress or accessories) 170 kg
Total weight authorized during moving 420 kg
Maximum temperature of the applied parts at 40°C 56.5° C
Unweighted peak acoustic pressure levels <120 dB
Weighted maximum measured acoustic pressure 42 dBA
a. These are average values, which may vary according to manufacturing tolerances.
b. Dimensions in mm.
c. Maximum incline in relation to the sleep surface
d. SWL 250 kg / the maximum patient weight varies according to the mattress and accessories used
- 185 kg as per EN60601-2-52 (intensive and acute care)
- 215 kg as per EN60601-2-52 (ambulatory care)
Features Value

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hillrom Accella 900 and is the answer not in the manual?

Hillrom Accella 900 Specifications

General IconGeneral
BrandHillrom
ModelAccella 900
CategoryMedical Equipment
LanguageEnglish