Specifications
8 Specifications
125
These specifications apply to the 3196 Power Quality Analyzer
Specifications 8
Environmental & Safety Specifications
Operating environment
Indoors, altitude up to 2000 m (6562-ft.)
Storage temperature
and humidity
-20°C to 50°C (-4°F to 122°F), 80% RH or less (non-conden-
sating)
Operating temperature
and humidity
0°C to 40°C (32°F to 104°F), 80% RH or less (non-condensat-
ing)
(Battery pack and PC Card operation is not supported between
-10°C and 0°C (14°F to 32°F), and LCD operation and accura-
cy are not guaranteed in this range)
Maximum input voltage
(50/ 60 Hz)
Voltage inputs: 780 Vrms, 1103 V peak
Current inputs: 1.7 Vrms, 2.4 V peak
Maximum rated voltage
to earth
Voltage input terminals: 600 Vrms (50/60 Hz)
Dielectric strength
(50/60 Hz for 1 min.)
5.55 kVrms for one minute (current sensitivity 1 mA)
Between voltage and clamp input terminals, between voltage
input terminals and instrument case, between voltage input ter-
minals U1 to U3 and U4
Enclosure protection
IP30 (per EN60529)
Applicable Standards
Safety EN61010-1:2001
Voltage Input: Pollution Degree 2,
Measurement Category III (anticipated transient over-
voltage 6000V)
EMC EN61326:1997+A1:1998+A2:2001 Class A
EN61000-3-2:2000
EN61000-3-3:1995+A1:2001