EasyManua.ls Logo

Hioki 3196 - Single-Phase System Connections

Hioki 3196
158 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1
2
3
4
5
6
7
8
9
10
11
Connecting to Lines to be Measured
4.2 Connection Methods
47
4.2.1 Single-Phase System Connections
Source side
L
O
A
D
Load side
A
N
G
Black
Red
S
O
U
R
C
E
Face the arrow toward the Load
A: Line, N: Neutral, G: Ground
Black
Gray
U- U+ U- U+ U- U+
U- U+
U INPUT
I INPUT
CH1 CH2 CH3 CH4CH1 CH2 CH3 CH4
Ground line mea
-
surement, etc.
Single-phase 2-wire (1P2W)
U1=100 V
I1=15 A
U4=10 V
I4=1.5 A
Source side
S
O
U
R
C
E
L
O
A
D
Load side
A
N
B
Black
Red
Yellow
Black
G
A: Line, N: Neutral, G: Ground
Black
Gray
U- U+ U- U+ U- U+
U- U+
U INPUT
I INPUT
CH1 CH2 CH3 CH4CH1 CH2 CH3 CH4
Ground line mea
-
surement, etc.
Face the arrow toward the Load
Single-phase 3-wire (1P3W)
U1=U2=100 V
I1=I2=15 A
U4=10 V
I4=1.5 A