3.4 Connecting the AC Adapter
32
Quick Start Manual
3.4 Connecting the AC Adapter
Use only the supplied Model 9458 AC ADAPTER. AC
adapter input voltage range is 100 to 240 VAC (with ±10%
stability) at 50/60 Hz. To avoid electrical hazards and dam-
age to the instrument, do not apply voltage outside of this
range.
Turn the instrument off before connecting the AC adapter
to the instrument and to AC power.
To avoid electric shock and ensure safe operation, connect
the power cord to a grounded (3-contact) outlet.
AC adapter operation with the 9459 BATTERY PACK installed
protects against loss of data when power is turned off.
Note that if the battery pack is not installed when operating with
the 9458 AC ADAPTER, a power interruption longer than 0.5 cycle
will abort measurement and cause the instrument to turn off.
When the AC adapter and battery pack are used together, the
AC adapter has priority.
1. Connect the power cord to
the AC adapter.
2. Connect the AC adapter
output plug to the instru-
ment.
3. Plug the power cord into
the AC mains outlet.
AC Adapter
Insert the plug into the outlet
(example)
Power Cord
1
2
3
Connecting the AC Adapter to the Instrument
Insert with the
arrow pointing
upward.