EasyManua.ls Logo

Hioki 3196 - 9 Maintenance and Service; Cleaning and Storage

Hioki 3196
158 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1
2
3
4
5
6
7
8
9
10
11
Maintenance and Service
9.1 Cleaning and Storage
137
Cleaning
Storage
Storage temperature and humidity should be kept between -20
and 50°C, at less than 80% RH.
Do not store or use the product where it could be exposed to
direct sunlight, high temperature or humidity, or condensation.
Under such conditions, the product may be damaged and insula-
tion may deteriorate so that it no longer meets specifications.
We recommend removing the battery pack when storing the
instrument. However, the battery pack is subject to self-dis-
charge. So when storing the battery pack for a long time, charge
the battery at least once every two months.
When storing the instrument for a long time (six months or more),
the specifications are no longer guaranteed. Therefore, before
use, have the instrument recalibrated.
Maintenance and Service 9
9.1 Cleaning and Storage
To clean the product, wipe it gently with a soft cloth moistened
with water or mild detergent. Never use solvents such as ben-
zene, alcohol, acetone, ether, ketones, thinners or gasoline, as
they can deform and discolor the case.
Wipe the LCD gently with a soft, dry cloth.
Measurements are degraded by dirt on the mating surfaces of
the clamp-on sensor, so keep the surfaces clean by gently wip-
ing with a soft cloth.