Appendix
144
Quick Start Manual
Changes on the Har-
monic screen do not
appear on the TIME
PLOT screen.
The operating settings do not record harmonics.
The unbalance factor
does not appears in the
TIME PLOT Change in
RMS Value screen.
The wiring setting is 1P2W, 1P3W or 3P3W2M.
Data Analysis problems
What is the display-
able period for analy-
sis?
14 cycles can be displayed at 50 Hz, or 16 cycles at 60
Hz.
None of the data that
was supposed to be
saved to the PC card
was saved.
Or, the intended data
was not saved.
Text save options must be selected.
The error message
"File not loaded com-
pletely." appears when
you try to open text
data with Excel.
Copy the data into a text editor, then split it into two files
and use the two parts separately.
We recommend that the number of text save options
selected in the 3196 SYSTEM screen be limited to no
more than 250 items.
Rec Data
Select either P&Harm or ALL DATA.
RECORDING
Wiring
Select 3P3W3M or 3P4W.
MEASURE
PC-CARD
SAVE
TEXT Select
Select the text
save options.
TEXT FILE
ave the data to the PC card again ([ANALYZING]
ppears while the data is being saved).
Auto Save
TEXT
RECORDING
Select items
Pre-measurement settings (condition in which [SET-
TING] appears)