2.2 Screen Names and Display Elements
28
Quick Start Manual
*5: SYSTEM Settings Indicators
*6: Number of repeated recording operations
Measurement Line type
Set on the SYSTEM screen
Voltage/Current Range
Red Indication: means out of
range
In this case, increase the
range setting.
Frequency of Measured
Line
Red Indication: means the
measurement frequency is dif-
ferent from the line frequency
CH4 Voltage mea-
surement function
(AC/ DC/OFF)
PLL Sync Frequency source
Indicates red when no input is applied
at the selected source.
PT/CT Ratio
[SC] (scaling) Appears when th
PT or CT ratio has been set.
(Not displayed): Displayed when
the PT and CT ratios are both 1.
When the repeated recording function is set, the number of
repeated recording operations currently set will be indicated.
Repeated recording operations can be conducted up to 99 days
at one-day measuring intervals, and up to 99 weeks at one-week
measuring intervals.
The measured data file of repeated recording is saved as a sepa
rate binary file for each one-day or one-week period on the PC
card.