Measurement Preparations
3.7 Using a Clamp-On Sensor (Option)
39
9694 CLAMP ON SENSOR
Accuracy of the combination of the 9660/9661/9669/9694 and the
3196:
Example: Using the 9661, 50 A range setting (3196), when measuring 50 Arms
Error of 9661 + Error of 3196 = Total error
Error of 9661 = 50 A(measurement) x 0.3% rdg. + 500 A(rating) x 0.01% f.s.
= (50 x 0.3/100) + (500 x 0.01/100)
= 0.15 + 0.05
= 0.20 A
Error of 3196 = 50 A (measurement)x 0.2% rdg. + 50 A (range) x 0.1% f.s.
= (50 x 0.2/100) + (50 x 0.1/100)
= 0.1 + 0.05
= 0.15 A
Total error = 0.20 A + 0.15 A
= ±0.35 A
9669 CLAMP ON SENSOR
Primary current rating 5 A AC
Output voltage 10 mVAC/A
Maximum input current 50 A continuous (at 45 to 66 Hz , 50°C)
Amplitude accuracy ±0.3%rdg. ±0.02%f.s.(f.s.=5 A, 45 to 66 Hz)
Phase accuracy ±2° or less (45 Hz to 5 kHz)
Amplitude frequency
characteristic
±1% accuracy or better from 40 Hz to 5 kHz
Maximum rated voltage
to earth
300 Vrms
Measurable conductor
diameter
15 mm max.
Operating temperature and
humidity
0 to 50°C (32 to 122°F), 80% RH or less
Primary current rating 1000 A AC
Output voltage 0.5 mV AC/A
Maximum input current 1000 A continuous (at 45 to 66 Hz , 50
°C)
Amplitude accuracy ±1.0% rdg. ±0.01% f.s. (f.s.=1000 A , 45 to 66 Hz)
Phase accuracy ±1
° or less (45 Hz to 5 kHz)
Amplitude frequency
characteristic
±2% accuracy or better from 40 Hz to 5 kHz
Maximum rated voltage
to earth
600 Vrms
Measurable conductor
diameter
55 mm max.
Operating temperature and
humidity
0 to 50
°C (32 to 122°F), 80% RH or less