11
Measurement Method
6.3 Harmonic Fluctuation
71
2.
Making 3196 settings (measurement settings)
MEASURE
Make the following settings.
Setting Item Selections
123ch/ Wiring
3P3W3M
123ch/ Clamp
10mV/A 9694
123ch/ U Range
150V
123ch/ PT Ratio
60
123ch/ I Range
50 A
123ch/ CT Ratio
60
UReference
110V
Frequency
50 Hz
PLL Source
U1
U CalcType
LINE-LINE
Harm Calc
LEVEL
THD Calc
THD_F
EN50160
OFF
Select item
Select from
pull-down
menu
Confirm
3.
Connecting to the lines to be measured
Use the same method as with three-phase checking. (page 59)
4.
Checking the connections
Use the same method as with three-phase checking. (page 60)